Our enterprise has been operating in the field of development of instrumentation and testing equipment for space industry for over 50 years. Basing on the experience of designing electronic products we have developed a multiplex strain measuring system, which is currently in operation on the strength test stand of aviation structures at ANTK «Antonov».

Automated information strain measurement system (/ATIS)

atis.jpg It is designed for examining and identifying strain-deformation state of structures with digital processing of measurement results.

Specification

Quantity of control points per measurement channel (with one active resistance strain gage) 100
Quantity of measurement channels (customized) from 1 to 128
Total cycle time of polling, measuring and sending the results into PC by all operating channels, s up to 8
Nominal value of the least significant bit of the measured value code, microohm /Ohm up to 4
Main measurement reduced error, % up to 0,2
Operating temperature range, from +5 to +35
Communication line length
From resistance strain gages to switch boards, m up to 10
From switch boards to the measuring device, m up to 40
From the measuring device to the PC, m up to 100
Dimensions
measuring device , mm 480x450x355
switch board for 100 measurement points, mm 390x170x115
  • The system provides diagnostics of failures to the plug-in unit, as well as diagnostics of sensor faults of the item under control.
  • Service Life Limit at least 5000 hours over the specified life of 10 years.
  • The system has a certificate of the state metrological attestation. The system may be customized according to the individual design. We provide service support over the whole operation period.

Automated multiplex information strain measurement system (-M/ATIS-M)

atism.jpgIt is designed for examining strain-deformation state of full-scale structures under the impact of static and dynamic loads with digital processing of measurement results.

Main features for configuring static tests:

Quantity of control points per measurement channel
Connecting resistance strain gages under 1/4 bridge scheme 100
Connecting resistance strain gages under 1/2 bridge or a bridge scheme 50
Total cycle time of polling, measuring and sending the results into PC by all operating channels, s up to 8
Nominal value of the least significant bit of the measured value code, microohm /Ohm up to 4
Main measurement reduced error, % up to 0,2
Communication line length
From resistance strain gages to switch boards, m up to 10
From switch boards to the measuring device, m up to 40

Main features for configuring dynamic tests:

Maximum quantity of measurement channels while connecting resistance strain gages under 1/2 bridge or a bridge scheme and structural loading frequency
from 0 to 50 Hz 4
from 0 to 200 Hz 2
from 0 to 1000 Hz 1
Recording range of structural loading frequencies
Operating preliminary data accumulation, Hz 0 1000
Operating online data transmission, Hz from 0 to 100
Main measurement reduced error (statistical), % up to 2
Maximum time of preliminary data accumulation, s up to 3600
Communication line length from resistance strain gages to the measuring device
Connecting resistance strain gages under 1/2 bridge scheme, m up to 10
Connecting resistance strain gages under the bridge scheme, m up to 70
System operation temperature range, from +5 to +35
Communication line length from the measuring device to the PC, m up to 100
  • The system provides diagnostics of failures to the plug-in unit, as well as diagnostics of sensor faults of the item under control.
  • The basic model may include up to 4 measurement modules of any configuration.
  • The system may be customized according to the individual design with increased quantity of measurement channels and control points.
  • Service Life Limit at least 5000 hours over the specified life of 10 years.